CAS ER JUNIOR

ER JUNIOR / ER PLUS

PUT THE SCALE INTO CAL (HOLD CAL BUTTON) AND TURN ON SCALE

PRESS THE – 0 – KEY UNTIL DISPLAYS CAL 3

PRESS THE – T -SCALE DISPLAYS ZERO

PRESS THE – T -TO TAKE UNLOAD READING

LOAD 1/3 CAPS I.E. 15 KG SCALE LOAD 5KG AND PRESS -T

TURN OFF SCALE

 

ER JUNIOR—EASY PLU PROGRAM

ENTER THE PRICE

PRESS AND HOLD THE * KEY

PRESS THE PLU POSITION (M1—M2)

PRESS T TO WND OR FOR TEXT PG 30 ASCII CODE

 

ER PLUS/ER-M

PLU PROGRAM KEY

PLU INDIRECT KEY

 

EASY PLU PROGRAM—DIRECT KEYS & INDIRECT

ENTER PRICE

PRESS (DIRECT PRESS PLU POSITION M1—M7) OR ENTER PLU NUMBER

PRESS T TO END

  

ER JNR 15 CALIBRATION

REMOVE CAL STICKER UNDER THE BASE OF SCALE

HOLD CAL SWITCH IN WHILST YOU TURN SCALE ON

SCALE WILL FLASH UP CAL 1

USING THE BUTTON WITH THE -0-   PRESS IT TWICE TO SHOW CAL 3

THEN USING THE -T – PRESS TWICE AND WAIT FOR THE SCALE TO COUNT AND SHOW IDUP

PUT 5 KG ON SCALE AND PRESS -T – AGAIN AND WAIT FOR SCALE TO COUNT AND DISPLAY SHOWS FULL.

PUT 15KG ON AND PRESS – T – UNTIL IDDN IS DISPLAYED REMOVE WEIGHT DOWN TO 5KG AND PRESS -T- SCALE WILL THEN SHOW CAL 3 REMOVE THE WEIGHTS THEN TURN THE SCALE OFF

THEN TURN THE SCALE BACK ON TO RETURN TO NORMAL OPERATION   