CAS AP-LP1 SW-BW

AP

SLIDE CAL SWITCH ACROSS AND TURN ON SCALE

TYPE C3 THE DISPLAY SHOWS ULOAD

WITH EMPTY PLATE PRESS C

DISPLAY SHOWS LOAD

LOAD CAPACITY AND PRESS C

DISPLAY SHOWS END

PRESS C—0 DISPLAY SHOWS Err 2

SLIDE CAL SWITCH BACK TO NORMAL

 

LP1 

SLIDE CAL SWITCH ACROSS AND TURN ON SCALE

TYPE 3 THE DISPLAY SHOWS ULOAD

WITH EMPTY PLATE PRESS PRT

DISPLAY SHOWS LOAD

LOAD CAPACITY AND PRESS PRT

DISPAY SHOWS END

TURN SCALE OFF, RETURN CAL SWITCH BACK TO NORMAL AND TURN SCALE ON

 

CAS SW—SINGLE RANGE

HOLD IN CAL BUTTON AND TURN ON SCALE

PRESS TARE UNTIL DISPLAY SHOWS C3

PRESS ZERO , DISPLAY SHOWS ULOAD

WITH EMPTY PLATE PRESS ZERO

DISPLAY SHOWS LOAD

LOAD CAPACITY AND PRESS ZERO

DISPLAY SHOWS END

PRESS TARE UNTIL DISPLAY SHOWS C-0

PRESS ZERO

 

SW—NEW STYLE—DUAL RANGE

PUT SCALE INTO CAL (HOLD IN CAL BUTTON) AND TURN SCALE ON

PRESS – T – UNTIL DISPLAYS CAL 3

PRESS – O -TWICE FOR EMPTY READING

LOAD HALF MAX CAPACITY (5KG ONLY LOAD 2KG) AND PRESS- 0         

LOAD MAX CAPACITY AND PRESS – 0          

LOAD HALF CAPACITY (5KG ONLY LOAD 2KG) AND PRESS -0         

TURN OFF

 

SW NEW DUAL

TO CHANGE WEIGHING MODE TO LB/OZ OR KG/G PRESS AND HOLD -T         

AND TOGGLE USING H

 

AD—BW—SW

PUT INTO CAL

PRESS ON THEN TARE THEN ENTER

UNLOAD

PRESS ON/OFF

LOAD CAPACITY PRESS ON/OFF

SHOWS END

 

(AD & BW) TURN OFF AND ON

(SW) PRESS ZERO